HAMILTON-T1 for ventilator-assisted preoxygenation

06.06.2018
Author: Clinical Experts Group, Hamilton Medical, Reviewer: N/A

The use of rapid sequence intubation (RSI) in the Emergency Department is often associated with complications, including serious oxygen desaturation.

The use of CPAP and/or NIV for preoxygenation may be particularly effective in reducing the risk of hypoxemia during intubation. A recent study from Australia now describes use of the HAMILTON-T1 to apply a combination of NIV and pressure-controlled ventilation in high-risk patients (1).

The authors' VAPOX technique (ventilator-assisted preoxygenation) is made possible by the HAMILTON-T1’s biphasic design and open valve system, which enable the application of NIV for preoxygenation followed by mechanical ventilation after intubation, all with the same breathing circuit.

After completing the preop tests and connecting the oxygen and power supplies, the HAMILTON-T1 is set up in NIV-ST mode with the following settings:

  • RR: 6–8 breaths/min
  • Psupport: 10 cmH2O
  • PEEP: 5 cmH2O
  • Oxygen: 100%
  • ETS (expiratory trigger sensitivity): 50%
  • Inspiratory flow trigger: 2 l/min
  • Ti: 2 s
  • P-ramp: 50 ms

A nasal cannula is applied to the patient with an oxygen flow rate of 15 l/min and then a face mask is applied. The ventilator is started and preoxygenation applied for 3 minutes.

As soon as apnea ensues, the ventilator transitions to PCV+. 

Intubation starts after at least 45 sec, with nasal oxygenation continuing until the ETT is in place. 

With the tube successfully positioned, the ventilator can be switched to a conventional ventilation mode.

In this series of patients, there was no occurrence of hypoxia. The two patients with the lowest oxygen saturation at the outset (77% and 79%) increased to 100% and 94% post VAPOX, respectively, with values post intubation at 93% and 92%, respectively. These results showed VAPOX to be both safe and effective.

Using the HAMILTON-T1, this promising technique can successfully be applied to minimize hypoxemia during RSI.

Relevant devices: HAMILTON-T1

References

  1. Grant S, Khan F, Keijzers G, Shirran M, Marneros L. Ventilator-assisted preoxygenation: Protocol for combining non-invasive ventilation and apnoeic oxygenation using a portable ventilator. Emerg Med Australas. 2016 Feb;28(1):67-72 (abstract only).
rsi, rapid sequence oxygenation, preoxygenation, t1, hamilton-t1, niv, cpap, hypoxemia, vapox, pressure controlled, intubation
Back

Date of Printing: 16.08.2018
Disclaimer:
The content of this Knowledge Base is intended for informational purposes only. Hamilton Medical AG provides no warranty with respect to the information contained in this Knowledge Base and reliance on any part of this information is solely at your own risk. For detailed instructions on operating your Hamilton Medical device, please refer to the official Hamilton Medical Operator’s Manual for the respective device.