Test your IntelliVence (Vol 15 Issue 6)

07.01.2019
Author: Simon Franz, Reviewer: Paul Garbarini, David Grooms
Last change: 07.01.2019, Revision , Additional image

This intubated patient is being ventilated in PSIMV+ mode. What kind of asynchrony is visible in the breath marked by the cursor?

 

 

 

 

 

 

 

Answer: What you can see here is delayed cycling. The ventilator’s mechanical inspiratory time exceeds the patient’s neural inspiratory time. In this case, delayed cycling is indicated by distortions in both the flow and pressure waveforms at mid-inspiration (~ 7 sec.). The flow curve shows a change in slope towards the baseline, indicating an abrupt relaxation of the inspiratory muscles. At the same time, the pressure curve shows a slight peak.

In this patient, the delayed cycling occurred during a mandatory breath.

Possible solutions are:

  • Shorten the set TI to match the patient’s neural TI.
  • Activate Psync to allow the patient to control his TI via flow-cycling at every breath. The set rate will function as a backup rate.
Delayed cycling, cycling, flow curve, asynchrony, PSIMV
Back

Date of Printing: 17.11.2019
Disclaimer:
The content of this newsletter is for informational purposes only and is not intended to be a substitute for professional training or for standard treatment guidelines in your facility. Any recommendations made in this newsletter with respect to clinical practice or the use of specific products, technology or therapies represent the personal opinion of the author only, and may not be considered as official recommendations made by Hamilton Medical AG. Hamilton Medical AG provides no warranty with respect to the information contained in this newsletter and reliance on any part of this information is solely at your own risk.