Test your IntelliVence (Vol 16 Issue 2)

23.04.2019
Author: Jean-Michel Arnal, Senior Intensivist, Ste Musse Hospital, Toulon, France, Reviewer: David Grooms

The volumetric capnography loop displayed is from a passive patient ventilated in PC mode. What does its shape tell us?

Answer:

Airway dead space seems to be in the normal range; it is calculated by the ventilator to be 114 ml.

The slope of Phase II is flatter than usual, suggesting an increase in airway resistance.

The slope of Phase III is increased, suggesting considerable heterogeneity in the lung with alveoli having a widely varying ventilation/perfusion ratio.

 

PetCO2 is low with a large gradient between PaCO2 and PetCO2, which confirms the severe ventilation/perfusion mismatching.

This is a patient with COPD with pneumonia ventilated with PEEP of 12 cmH2O. The PEEP setting may not have been appropriate in this case.

volumetric capnography, airway deadspace, PEEP, ventilation perfusion ratio, airway resistance, ventilation perfusion mismatching
Back

Date of Printing: 17.11.2019
Disclaimer:
The content of this newsletter is for informational purposes only and is not intended to be a substitute for professional training or for standard treatment guidelines in your facility. Any recommendations made in this newsletter with respect to clinical practice or the use of specific products, technology or therapies represent the personal opinion of the author only, and may not be considered as official recommendations made by Hamilton Medical AG. Hamilton Medical AG provides no warranty with respect to the information contained in this newsletter and reliance on any part of this information is solely at your own risk.