Can I measure work of breathing for patient and ventilator?

09.04.2020
Author: Munir Karjaghli, Reviewer: Bernhard Schmitt, Thomas Reimer, Matthias Himmelstoss, Kaouther Saihi

Work of breathing (WOB) represents the integral of the product of volume and pressure. It represents the energy associated with a given VT at a given pressure (spontaneous, mechanical, or both).

The airway pressure is the pressure of the whole respiratory system (lungs plus chest wall), the transpulmonary pressure is the pressure needed to distend the lung parenchyma, and the pleural pressure is the pressure needed to distend the chest wall.

In paralyzed patients with mechanical ventilation, plots of airway pressure versus VT indicate the total amount of work needed to inflate the respiratory system. This represents the work done by the ventilator on the whole respiratory system and the ETT, NOT the amount of work performed by the respiratory muscles.

Calculation of a patient’s WOB may be useful for understanding weaning failure, for helping in titrating ventilator support, and for evaluating the effects of different ventilation modes, the effects of therapeutic interventions, and the influence of ventilator performance (triggering, flow delivery, etc.).

For healthy people, the average total WOB ranges from 0.3 to 0.6 J/L. Patients with severe obstructive or restrictive lung disease “work” at levels two to three times this normal value at rest, with marked increases in work at higher minute ventilation (1). How much work a patient can tolerate before the respiratory muscles tire is unclear. Maintaining adequate spontaneous ventilation is impossible in many patients when the workload exceeds 1.5 J/L.

The pressure-time product is the integral of the pressure performed by the respiratory muscles during inspiration. It is an alternative to WOB and has some theoretical and practical advantages over WOB calculations. The pressure-time product (PTP) is the area encompassed by the esophageal or airway pressure time tracing during inspiration.

In principle, the HAMILTON-G5/S1 ventilator measures the WOB for the patient only; this is valid for Paw and Paux (whichever is selected). WOB is airway pressure integrated over inspiratory volume until the pressure exceeds the PEEP/CPAP level. In the dynamic pressure/volume loop, WOB is the area to the left of PEEP/CPAP (as shown in Figure 1).

The PTP is the measured pressure drop required to trigger the breath multiplied by the time interval until the PEEP/CPAP level is reached at the beginning of inspiration. The PTP on Hamilton Medical ventilators is the area between PEEP and Paw or Paux/Pes (whichever is selected; Paux/Pes only for HAMILTON-G5/S1 and HAMILTON-C6, respectively), as long as Paw/Paux (Pes) is below PEEP as shown in Figure 2. Normal values for the pressure-time product range between 60 and 150 cmH2O/second per minute.  

Figure 1
Figure 2

 

 

 

 

 

 

 

 

 

 

Relevant ventilators: All
Relevant software: HAMILTON-C1/T1/MR1 SW v2.2.4 and lower; HAMILTON-C2 SW v 2.2.5 and lower; HAMILTON-C3 SW v 2.0.5 and lower; HAMILTON-G5/S1 SW v 2.81 and lower

References

  1. Layon J, Banner MJ, Kirby RR, et al: Partially and totally unloading respiratory muscles based on real-time measurements. Chest 106:1835–1842A, 1994.
WOB, work of breathing, load, airway pressure, pleural pressure, transpulmonary pressure, pressure time product, Paw, Pes, Paux
Back

Date of Printing: 13.08.2020
Disclaimer:
The content of this Knowledge Base is intended for informational purposes only. Hamilton Medical AG provides no warranty with respect to the information contained in this Knowledge Base and reliance on any part of this information is solely at your own risk. For detailed instructions on operating your Hamilton Medical device, please refer to the official Hamilton Medical Operator’s Manual for the respective device.