Bedside tip: Using volumetric capnography for setting PEEP

23.04.2019
Author: Jean-Michel Arnal, Senior Intensivist, Ste Musse Hospital, Toulon, France, Reviewer: David Grooms

PEEP is used to keep the lung aerated and prevent lung collapse at the end of expiration. However, PEEP may over-distend the normally aerated lung and impair lung perfusion. Therefore, any change in PEEP may affect the overall ventilation/perfusion ratio in an unpredictable way.

Volumetric capnography measures the volume of CO2 exhaled breath-by-breath (VeCO2). After a change in PEEP, assuming that the cardiovascular function and tidal volume are stable, an increase in VeCO2 means that the overall ventilation/perfusion ratio has improved. Conversely, a decrease in VeCO2 means that the ventilation/perfusion ratio is worsening. VeCO2 changes rapidly and returns to the baseline after a few minutes.

The limitation to this method is that the clinician monitors rapid changes to the ventilation/perfusion ratio, such as those due to lung over-distension and lung-perfusion impairment or improvement.

After a change in PEEP, recruitment or derecruitment may take a longer time to occur and cannot be assessed by this method.

Watch the video to see a demonstration during ventilation with a Hamilton Medical ventilator. 

Reference:

  • Detection of optimal PEEP for equal distribution of tidal volume by volumetric capnography and electrical impedance tomography during decreasing levels of PEEP in post cardiac-surgery patients. P. Blankman, A. Shono, B. J. M. Hermans, T. Wesselius, D. Hasan, D. Gommers. British Journal of Anaesthesia, 116 (6): 862–9 (2016)

Related Articles

volumetric capnography, setting PEEP, PEEP, ventilation perfusion ration, recruitment, VeCO2
Back

Date of Printing: 15.11.2019
Disclaimer:
The content of this newsletter is for informational purposes only and is not intended to be a substitute for professional training or for standard treatment guidelines in your facility. Any recommendations made in this newsletter with respect to clinical practice or the use of specific products, technology or therapies represent the personal opinion of the author only, and may not be considered as official recommendations made by Hamilton Medical AG. Hamilton Medical AG provides no warranty with respect to the information contained in this newsletter and reliance on any part of this information is solely at your own risk.