Author: Dr. med. Jean-Michel Arnal, Senior Intensivist, Hopital Sainte Musse, Toulon, France, Reviewer: Paul Garbarini, David Grooms

This patient is passive and ventilated in pressure-control mode. Why is the flow curve shaped this way?

 

 

 

 

 

 

 

 

Answer: This shape indicates a severe flow limitation in a COPD patient. A flow limitation usually occurs during expiration and is represented by bicompartmental expiration. The initial spike of expiratory flow corresponds with a dynamic compression of gas in the central airways. Thereafter, the flow decreases rapidly because it is limited at the choke point. In the case of severe bronchospasm, the inspiratory flow may display the same shape with an initial fast compartment followed by a slow compartment due to the flow limitation at the choke point.

References

  • Lourens MS, van den Berg B, Verbraak AF, Hoogsteden HC, Bogaard JM (2001) Effect of series of resistance levels on flow limitation in mechanically ventilated COPD patients. Respir Physiol 127:39–52
  • Valta P, Corbeil C, Lavoie A, Campodonico R, Koulouris NG, Chasse´M, Braidy J, Milic-Emili J. Detection of expiratory flow limitation during mechanical ventilation. Am J Respir Crit Care Med 1994;150:1311–1317.
pressure control mode, flow limitation, COPD, expiratory flow, bicompartmental expiration, choke point, brochospasm
Back

Date of Printing: 10.12.2018
Disclaimer:
The content of this newsletter is for informational purposes only and is not intended to be a substitute for professional training or for standard treatment guidelines in your facility. Any recommendations made in this newsletter with respect to clinical practice or the use of specific products, technology or therapies represent the personal opinion of the author only, and may not be considered as official recommendations made by Hamilton Medical AG. Hamilton Medical AG provides no warranty with respect to the information contained in this newsletter and reliance on any part of this information is solely at your own risk.