This article gives you an overview of possible measures that may be taken to ensure protection against internal contamination of the ventilators, as well as protection of patients and clinical staff.
As the novel coronavirus (2019-nCoV) continues to spread and has now been declared an international emergency by the WHO, questions regarding the safe use of Hamilton Medical ventilators on patients with highly infectious diseases are becoming more frequent.
We therefore recommend implementing the following measures to avoid contamination:
- Follow the instructions for use of the ventilator and consider the WHO guidelines. (1, 2)
- Use an inspiratory bacterial and viral filter to assure non-contamination of the internal ventilator gas path.
- Protect the expiratory valve with a hydrophobic bacterial and viral filter.
- For active humidification, as with a HAMILTON-H900 humidifier, use a bacterial and viral filter on the inspiratory and expiratory port of the ventilator (hydrophobic version).
- For passive humidification, use a bacterial and viral HME/HMEF filter between the proximal flow sensor and the patient to protect the airway against contamination. Be aware of changes in anatomical dead space and airway resistance, and exchange the filters regularly.
- Contamination of the flow sensor tube connectors is avoided due to permanent rinse flow through the flow sensor tubing towards the patient.
- The Stand-by function can be used prior to disconnecting the ventilator from the patient to avoid mucus dispersion from the circuit. This decision should be made by the responsible clinician based on the situation of the individual patient.
- Use single-use consumables such as breathing circuits, flow sensor, airway adapters, expiratory valves, and filters to minimize the risk of cross-contamination when the ventilator has been cleaned and set up for a new patient.
- Disinfect the outer surfaces of the ventilators during ventilation or after treatment of a patient with a registered hospital disinfectant. Consult the hygiene specialist in your facility regarding the appropriate disinfectant and follow the instructions for use of the manufacturer, particularly with respect to contact time.
- For suctioning, use a closed inline suction system only.
- Reduce the need for user interaction with the ventilator by using Hamilton Medical’s INTELLiVENT-ASV mode for intubated patients. INTELLiVENT-ASV continuously adapts ventilation to the patient condition and requires fewer interactions by clinicians. (3, 4)
- All turbine-driven Hamilton Medical ventilators (HAMILTON-C6*/C3/C2/C1/T1/MR1) are equipped with high-grade HEPA filters to keep the interior airway free of contamination. There is no need to change the HEPA filters more frequently than indicated in your regular maintenance plan.
Make sure that all clinical staff involved in the handling of the ventilator are informed about the above-mentioned measures.
This information is available for download below.
* Not available in all markets
References
- Infection prevention and control of epidemic- and pandemic-prone acute respiratory infections in health care. World Health Organization, Geneva, 2014.
- Infection prevention and control during health care when novel coronavirus (nCoV) infection is suspected Interim guidance. World Health Organization, Geneva, 2020. Licence: CC BY-NC-SA 3.0 IGO
- Beijers AJR, Intensive Care Med. 2014 May;40(5):752-3.
- Arnal, J.M.,Minerva Anestesiol, 2018. 84(1): p. 58-67.
Downloads
Produit | Catégorie | Sorte/Taille | Date/PN | Langue |
---|---|---|---|---|
![]() HEPA filter company statement |
PDF 80.1 KB |
2018-11-06
ELO20180811N |
EN |
|
![]() Safe use of Hamilton Medical ventilators for patients with highly infectious diseases |
PDF 44 KB |
2021-01-05
ELO20200306N |
EN |
|
![]() Sichere Verwendung von Hamilton Medical Beatmungsgeräten an Patienten mit hoch ansteckenden Krankheiten |
PDF 41.6 KB |
2021-01-05
ELO20200307N |
DE |
|
![]() Uso seguro de respiradores de Hamilton Medical en pacientes con enfermedades altamente infecciosas |
PDF 40.4 KB |
2021-01-05
ELO20200308N |
ES |
|
![]() Utilisation sans risque des ventilateurs Hamilton Medical sur des patients atteints de maladies extrêmement contagieuses |
PDF 41 KB |
2021-01-05
ELO20200310N |
FR |
|
![]() Utilizzo sicuro dei ventilatori Hamilton Medical su pazienti con malattie altamente infettive |
PDF 41 KB |
2021-01-05
ELO20200311N |
IT |
|
![]() 高感染性疾患に罹患している患者での Hamilton Medical人工呼吸器の安全な使用 |
PDF 408.7 KB |
2021-01-05
ELO20200313N |
JA |
|
![]() Uso seguro de respiradores Hamilton Medical em pacientes com doenças altamente infecciosas |
PDF 40.7 KB |
2021-01-05
ELO20200309N |
PT |
|
![]() Меры безопасности во время использования аппаратов ИВЛ Hamilton Medical для пациентов с острыми инфекционными заболеваниями |
PDF 151.3 KB |
2021-01-05
ELO20200312N |
RU |
|
![]() Hamilton Medical 哈美顿医疗公司呼吸机在 高传染性疾病病人中的安全使用 |
PDF 352.5 KB |
2021-01-05
ELO20200314N |
ZH |
Disclaimer:
The content of this Knowledge Base is intended for use by Hamilton Medical distribution partners for informational purposes only, and should not be copied, distributed or published in any way. While the information contained herein is believed to be accurate, it does not represent an official recommendation from Hamilton Medical AG, nor may it substitute an opinion, assessment or instructions provided by a trained healthcare professional.