Bedside Tip: How to measure esophageal pressure correctly

19.10.2018
Author: Jean-Michel Arnal, Senior Intensivist, Hopital Sainte Musse, Toulon, France, Reviewer: Paul Garbarini, David Grooms

A recent physiological study demonstrated that esophageal pressure estimates the pleural pressure at mid-thorax at all levels of PEEP. Therefore, an absolute measurement of esophageal pressure is useful for setting PEEP and monitoring transpulmonary pressure.

So how do we measure esophageal pressure correctly?

The esophageal balloon has to be positioned and inflated correctly, and placement then verified.

Positioning

The optimal position for the esophageal balloon is the lower third of the esophagus, at a distance of 35–45 cm from the nostrils. With a patient in semi-recumbent position, the empty balloon is first inserted into the stomach, which is situated around 50–60 cm from the nostrils. The balloon is inflated to a standard volume (1 ml for a Cooper Surgical catheter and 4 ml for a Nutrivent catheter). Gastric pressure displays a positive deflection during inspiration in both passive and spontaneously breathing patients. The gastric position is ascertained by gently applying manual epigastric compression, which shows an immediate increase in gastric pressure.

 

 

The esophageal catheter is then gently withdrawn while the balloon is still inflated, in order to position the balloon in the lower third of the esophagus. During the change from gastric to esophageal pressure, the baseline of the pressure waveform changes and cardiac oscillations appear.

Gastric pressure (left) and esophageal pressure (right)

 

 

 

 

 

 

 

 

 


Esophageal pressure deflections are positive during inspiration in passive patients, but negative in spontaneously breathing patients. If cardiac oscillations distort the esophageal pressure signal, you can withdraw the catheter by another 2–5 cm.

Passive patient (left) and spontaneously breathing patient (right)

 

 

 

 

 

 

 

 

 

 

 

 

Balloon inflation

Balloon inflation performed with a Nutrivent catheter

The volume of air for adequate inflation of the balloon should be individually titrated. This is only possible in passive patients. According to the method proposed by Mojoli et al (2016), the balloon is inflated from 0.5 to 3 ml in gradual steps of 0.5 ml for a Cooper Surgical catheter, and from 1 to 8 ml in steps of 1 ml for a Nutrivent catheter. During progressive inflation of the balloon, the baseline of esophageal pressure increases and the magnitude of the esophageal pressure deflection changes. The adequate inflation volume is the one associated with the largest deflection of esophageal pressure. If two different inflation volumes show the same magnitude of esophageal pressure deflection, the lowest inflation volume is selected.

 

 

 

Verification

Once the balloon has been positioned correctly in the esophagus and inflated, verification is performed by means of an occlusion test. The principle is to close the airways at end expiration to change the airway pressure, and then to verify that esophageal pressure changes by the same amount.

In passive patients, you can perform an end-expiratory occlusion. When the expiratory valve is closed, apply an external manual compression of the rib cage on both sides of the chest in order to see a positive deflection of airway and esophageal pressures. The magnitude of the increase in airway and esophageal pressure should be the same. In other words, the transpulmonary pressure should not change.

In active patients, the dynamic occlusion test also uses an end-expiratory occlusion. There is no need to press manually on the chest, as the patient will make a spontaneous inspiratory effort during the occlusion. The result is a negative deflection of airway and esophageal pressure. The magnitude of the decrease in airway and esophageal pressure should be the same, i.e., the transpulmonary pressure should not change.

Occlusion test in active patient (left) and passive patient (right)

 

 

 

 

 

 

 

 

 

 

If you wish to monitor esophageal pressure continuously, it is important to reassess the correct position and inflation volume.


References

Yoshida T, Amato MBP, Grieco DL, Chen L, Lima CAS, Roldan R, et al. Esophageal Manometry and Regional Transpulmonary Pressure in Lung Injury. Am J Respir Crit Care Med 2018;197(8):1018-1026 (abstract only).

Mojoli F, Iotti GA, Torriglia F, Pozzi M, Volta CA, Bianzina S, et al. In vivo calibration of esophageal pressure in the mechanically ventilated patient makes measurements reliable. Crit Care 2016;20:98.

Baydur A, Behrakis PK, Zin WA, Jaeger M, Milic-Emili J. A simple method for assessing the validity of the esophageal balloon technique. Am Rev Respir Dis 1982;126:788–791 (abstract only).

Akoumianaki E, Maggiore SM, Valenza F, Bellani G, Jubran A, Loring SH,et al.; PLUG Working Group (Acute Respiratory Failure Section of the European Society of Intensive Care Medicine). The application of esophageal pressure measurement in patients with respiratory failure. Am J Respir Crit Care Med 2014;189:520–531 (abstract only).

Mauri T, Yoshida T, Bellani G, Goligher EC, Carteaux G, Rittayamai N, et al; PLeUral pressure working Group (PLUG—Acute Respiratory Failure section of the European Society of Intensive Care Medicine). Esophageal and transpulmonary pressure in the clinical setting: meaning, usefulness and perspectives. Intensive Care Med 2016;42(9):1360-73 (abstract only).

Related Articles

esophageal pressure measurement, how to measure, positioning, esophageal balloon, esophageal catheter, inflation, balloon filling, verification, occlusion test
Back

Date of Printing: 10.12.2018
Disclaimer:
The content of this newsletter is for informational purposes only and is not intended to be a substitute for professional training or for standard treatment guidelines in your facility. Any recommendations made in this newsletter with respect to clinical practice or the use of specific products, technology or therapies represent the personal opinion of the author only, and may not be considered as official recommendations made by Hamilton Medical AG. Hamilton Medical AG provides no warranty with respect to the information contained in this newsletter and reliance on any part of this information is solely at your own risk.